3701:1-58-48. Calibration measurements of brachytherapy sources
-
(A) Before the first medical use of a brachytherapy source, a licensee shall have:
(1) Determined the source output or activity using a dosimetry system that meets the requirements of paragraph (A) of rule 3701:1-58-60 of the Administrative Code;
(2) Determined source positioning accuracy within applicators; and
(3) Used published protocols currently accepted by nationally recognized bodies to meet the requirements of paragraphs (A)(1) and (A)(2) of this rule.
(B) Instead of a licensee making its own measurements as required in paragraph (A) of this rule, the licensee may use measurements provided by the source manufacturer or by a calibration laboratory accredited by the "American Association of Physicists in Medicine" (AAPM) that are made in accordance with paragraph (A) of this rule.
(C) A licensee shall mathematically correct the outputs or activities determined in paragraph (A) of this rule for radioactive decay at intervals consistent with one percent radioactive decay.
(D) A licensee shall retain a record of each calibration in accordance with rule 3701:1-58-89 of the Administrative Code.
Five Year Review (FYR) Dates:
06/12/2015 and
06/01/2020
Promulgated
Under:
119.03
Statutory Authority:
3748.02,
3748.04
Rule Amplifies:
3748.04
Prior Effective Dates: 8/15/2005,
10/4/10
Prior History: (Effective:
10/04/2010
R.C.
119.032 review dates:
06/28/2010 and
07/12/2015
Promulgated
Under:
119.03
Statutory Authority:
3748.02,
3748.04
Rule Amplifies:
3748.04
Prior Effective Dates: 8/15/2005
)