Ohio Administrative Code (Last Updated: January 12, 2021) |
3701:1 Radiation Control |
Chapter3701:1-58. Medical Use of Radioactive Materials |
3701:1-58-50. Therapy-related computer systems for manual brachytherapy
-
The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
(A) The source-specific input parameters required by the dose calculation algorithm;
(B) The accuracy of dose, dwell time, and treatment time calculations at representative points;
(C) The accuracy of isodose plots and graphic displays; and
(D) The accuracy of the software used to determine sealed source positions from radiographic images.
Five Year Review (FYR) Dates:
06/12/2015 and
06/01/2020
Promulgated
Under:
119.03
Statutory Authority:
3748.02,
3748.04
Rule Amplifies:
3748.04
Prior Effective Dates:
8/15/2005
Prior History: (R.C.
119.032 review dates:
07/12/2010 and
07/12/2015
Promulgated
Under:
119.03
Statutory Authority:
3748.02,
3748.04
Rule Amplifies:
3748.04
Prior Effective Dates: 8/15/2005
)