Ohio Administrative Code (Last Updated: January 12, 2021) |
3701:1 Radiation Control |
Chapter3701:1-58. Medical Use of Radioactive Materials |
3701:1-58-95. Records of periodic spot-checks for teletherapy units
-
(A) A licensee shall retain a record of each periodic spot-check for teletherapy units required by rule 3701:1-58-64 of the Administrative Code for three years.
(B) The record must include:
(1) The date of the spot-check;
(2) The manufacturer's name, model number, and serial number of the teletherapy unit, source and instrument used to measure the output of the teletherapy unit;
(3) An assessment of timer linearity and constancy;
(4) The calculated on-off error;
(5) A determination of the coincidence of the radiation field and the field indicated by the light beam localizing device;
(6) The determined accuracy of each distance measuring and localization device;
(7) The difference between the anticipated output and the measured output;
(8) Notations indicating the operability of each entrance door electrical interlock, each electrical or mechanical stop, each source exposure indicator light, and the viewing and intercom system and doors; and
(9) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.
(C) A licensee shall retain a copy of the procedures required by paragraph (B) of rule 3701:1-58-64 of the Administrative Code until the licensee no longer possesses the teletherapy unit.
Five Year Review (FYR) Dates:
06/12/2015 and
06/01/2020
Promulgated
Under:
119.03
Statutory Authority:
3743.02.
3748.04
Rule Amplifies:
3748.04
Prior Effective Dates:
8/15/2005
Prior History: (R.C.
119.032 review dates:
07/12/2010 and
07/12/2015
Promulgated
Under:
119.03
Statutory Authority:
3748.02,
3748.04
Rule Amplifies:
3748.04
Prior Effective Dates: 8/15/2005
)