3701:1-58-89. Records of calibration measurements of brachytherapy sources  


Latest version.
  • (A) A licensee shall maintain a record of the calibrations of brachytherapy sources required by rule 3701:1-58-48 of the Administrative Code for three years after the last use of the source.

    (B) The record must include:

    (1) The date of the calibration;

    (2) The manufacturer's name, model number, and serial number for the source and the instruments used to calibrate the source;

    (3) The source output or activity;

    (4) The source positioning accuracy within the applicators; and

    (5) The name of the individual, the source manufacturer, or the calibration laboratory that performed the calibration.